Thank you for taking an interest in the various Voltage Standard components that we offer here on ASAP Military Parts, such as those listed with NSN 6625013616809. The particular item that you have taken an interest in is listed alongside the part number 5790A, having been manufactured by Fluke Corporation. With current availability for this product, we encourage you to utilize our online RFQ forms to quickly and easily request quotes for your comparisons. Within 15 minutes of us receiving a completed request and reviewing your provided details, a member of our staff will reach out to you to present a tailored solution.
Note : We Will Not Share Your Information To Any Third Parties.
To make the part procurement process as simple as possible, we choose to list all offerings with as many details as we can to ensure that you find exactly what you require with ease. For example, with the NSN 6625-01-361-6809, one can determine that part number 5790A belongs to FSC 6625 Electrical and Electronic Properties Measuring and Testing Instruments. Furthermore, its manufacturer Fluke Corporation specifically produced this part under CAGE Code 89536, allowing you to know the exact facility it traces back to. If you are ready to begin procurement for part number 5790A, send us a completed RFQ form today. You may also give us a call or email at any time, and we would be more than happy to assist you however we can as your strategic sourcing partner!
MRC | Criteria | Characteristic |
---|---|---|
BHXX | ELECTROMOTIVE FORCE PRODUCTION METHOD | ELECTRONIC |
BHYB | WARM-UP TIME FOR MAXIMUM ACCURACY IN MINUTES | 30.0 |
CRTL | CRITICALITY CODE JUSTIFICATION | SUPP |
AARB | TERMINAL TYPE | CONNECTOR AND BINDING POST |
APQB | UNIT TYPE | ENCASED |
ABMK | OVERALL WIDTH | 17.000 INCHES NOMINAL |
FAAZ | PHASE | SINGLE |
SR-5 | DESIGN CONTROL REFERENCE | 5790A/AF |
FEAT | SPECIAL FEATURES | FREQUENCY RANGE 10HZ TO 1MHZ; VOLTAGE RANGE 700UV TO 1000V; TRACEABLE TO NATIONAL STANDARDS; 24PPM TOTAL UNCERTAINTY; AUTOMATIC AC MEASUREMENT; COMPATIBLE WITH FLUKE A40 AND A40A CURRENT SHUNTS; POWER UP SELF TEST; 4 SETS OF INPUT TERMINALS,1 TYPE N AND 1 SET OF BINDING POSTS DEDICATED FOR AC MEASUREMENTS AND TRANSFER MODES. 2ND SET OF BINDING POST ARE FOR CURRENT SHUNTS;DIGITAL DISPLAY SHOWN IN PPM,PCT,VOLTS AND RATIO 1200V INPUT PROTECTION ON ALL RANGES:GPIB/IEEE488,RS-232C INTERFACES ARE INCLUDED; OPERATING TEMP 0 TO 50DEG C; OPERATING ALTITUDE 10000 FT; EMI/RFI COMPLIES FCC PART 15. SUBPART B,CLASS B,VDE 0871,CLASS B:ESD:EIA PN-1361,SURGE ANSI C62.41-1980 CAT ARELIABILITY: MIL-T-28800D PARA 3.13.3. WEIGHT: 53LBS; LINE VOLTAGES SELECTAB LE 100,110,115,120,200,220,230 OR 240VOLTS AC |
ABFY | OVERALL DEPTH | 24.800 INCHES NOMINAL |
SUPP | III SUPPLEMENTARY FEATURES | 5790A/AF IS CHECKED AND VERIFIED BY BY AEROSPACE GUIDENCE AND METROLOGY CENTER BEFORE BEING SHIPPED TO THE PMEL LABS |
BHXZ | OUTPUT ACCURACY RATING IN PERCENT | -0.010 TO 0.010 |
ACZB | FREQUENCY RATING | 47.0 HERTZ MINIMUM AND 63.0 HERTZ MAXIMUM |
SR-5 | THE MANUFACTURERS DATA | |
ACYN | AC VOLTAGE RATING | 115.0 VOLTS NOMINAL AND 220.0 VOLTS NOMINAL |
APCB | PORTABILITY | NONPORTABLE |
AARA | TERMINAL QUANTITY | 1 AND 5 |
ABKW | OVERALL HEIGHT | 7.000 INCHES NOMINAL |
SR-5 | MANUFACTURERS CODE | 89536 |